HomeProductsMobil GC 2
OIL ANALYSIS

Mobil GC 2

Portable gas chromatograph for precise on-site analysis of dissolved gases – to ASTM D3612 and IEC 60567.

Ageing processes such as oxidation or thermal breakdown steadily degrade the chemical and dielectric properties of the insulating oil. With the Mobil GC 2 you can react to faults early and reduce unplanned maintenance.

ASTM D3612IEC 60567Vacuum extraction
Mobil GC 2mobile
11 gases
+ total gas content
40 h
internal gas supply
19 kg
rugged Pelicase
0.1 ppm
resolution
Product video

Mobil GC 2 on video

More videos

Features

All benefits at a glance

Portable & flexible

Gas-in-oil analysis for condition assessment of insulating oils – on site at the transformer or in the lab.

Fully automatic

Fully automatic measurement with vacuum extraction – guided step by step via the display.

Standard-compliant

Testing of dissolved gases to ASTM D3612 and IEC 60567 – reliable and comparable.

High sensitivity

High sensitivity when analysing 11 gases – including direct analysis of the Buchholz gases.

40 h continuous operation

Internal gas supply for up to 40 hours of continuous operation – independent of the location.

One calibration

Calibration with a single calibration gas – suitable for all new and aged insulating oils.

Process

How the measurement works

In just a few steps from the sample port to a standard-compliant result – the display guides you through the entire process.

01

Take a sample

Extract the oil sample without contamination – e.g. with the OSM2 Pro.

02

Fit the syringe

Place the syringe on the port of the Mobil GC 2 and start the measurement.

03

Degas automatically

After degassing, the gas sample is transferred automatically into the GC.

04

Get the result

Measurement result incl. total gas content – insight into the gas saturation.

Expert System software

The Mobil GC 2 comes with the Expert System software. Apply various diagnostic methods or design your own method with individual gas concentrations as the basis for condition assessment.

RogersIECDuvalCustom method

Detection limits

Accuracy · ppm
H₂
1 ppm
CO
0,2 ppm
CO₂
0,2 ppm
CH₄
0,1 ppm
C₂H₂
0,1 ppm
C₂H₄
0,1 ppm
C₂H₆
0,1 ppm
C₃H₆
0,5 ppm
C₃H₈
0,5 ppm
O₂
30 ppm
N₂
30 ppm
Σ
Total gas

The more gases detected, the more information about the condition of the insulating oil – the Mobil GC 2 analyses the total gas content and 11 fault gases.

Technical details

Measurement technique
Gas-in-oil analysis with vacuum extraction and GC
Measuring range
Gas in oil – 0.1 – 100,000 ppm
Resolution
0.1 ppm
Calibration
Traceable calibration gas for all new and aged insulating oils
Sampling
Glass syringe 10 ml, gas-tight
Power supply
100 – 240 VAC, 50/60 Hz
Operating temperature
0 – 50 °C · 5 – 95 % rel. humidity, non-condensing
Dimensions
480 × 500 × 310 mm
Weight
19 kg
Housing
Pelicase
Internal gas supply
Argon and hydrogen, both 99.999 % purity

Scope of delivery

  • Expert System software
  • Degassing flange
  • Calibration object
  • Glass syringes
  • Filling system for internal gas supply
  • Calibration gas for approx. 250 calibrations
Full datasheet (PDF)

Interested in the Mobil GC 2?